When your testing requires a walk in rooms choose “TEMPO” controlled temperatures solution. Our test rooms feature automated process control, energy efficiency and a vast range of other features and options. Every “TEMPO” environmental room features equipment that reliably produces required conditions, structural integrity that keeps the room working properly through years of demanding test cycles, and measuring/monitoring equipment that precisely records all test data.

Standard Models

Inside S.S. 304 with mirror polish & outside mild steel powder coated.

GMP Models

Inside S.S. 316 with mirror polish & outside stainless steel 304 with the dull buff finish.

Specification

Temperature range

 2°C to 8°C

±0.5°C

±2°C

Construction

Light weight sandwich panel. Double walled construction with backside triple wall. 3″ thick PUF insulation ensures stable temperature with a reduction in energy consumption. The motorized blower at the backside of the chamber for recirculation of air to develop a unique airflow system which ensures maximum uniform temperature condition inside the chamber. The unique airflow assures quick recovery after door openings.

Single-phase, TEFC, F class insulation 1440 rpm, 230V. HIgh-quality SS tubular heaters are used for better heating. Steam injection system and reservoir tank with water level arrangement with indicator.

Features

  • Negligible maintenance – the sturdy design and long-lasting finish ensure minimum maintenance for years.
  • Minimal heat transfer: the tongue and groove edges of all panels are moulded to close tolerance, ensuring a tight fit between panels. These prevent condensation from forming on the foam and minimize heat transfer.
  • Quality accessories—hardware like hinges, handles, etc.—on panels from the best indigenous suppliers.

Safety Features

  • Audio visual alarms are used for temperature and humidity variations.
  • If the temperature increases above the 60°C the system will automatically cut off the dry heater, wet heater and the compressor.
  • The float switch is used to cut off the wet heater supply in case of the water level in the tank goes below.

Technical Specification

Controlling System

Microprocessor based PLC (up to 5000 readings) with 7″ touch screen colour HMI, which gives the direct display of process value and set the value of temperature and humidity inside the chamber. High speed ethernet based communication.

The refrigeration is provided by means of CFC free hermetically sealed Emerson Copland compressor for better cooling with eco-friendly refrigerant.  The cooling system is made up of split type so as one can place the compressor coupled with condenser unit in a well ventilated area to avoid hot air throw in the stability room. The inbuilt stand by refrigeration system is provided with auto change over the facility.

Optional Features

  • CFR 21 part 11 compliance software
  • Standby refrigeration and humidity system
  • Ethernet communication
  • Door access system
  • Data logger/scanner
  • Touch screen HMI
  • PLC-based control
  • Email alert
  • Mobile alert system
  • Safety alarms
Models & Optional Features
ModelsCapacity (litres)

Internal Dimensions

H × W × D (cm)

Trays
TI – 810 WIC5400200 × 136 × 20024
TI – 811 WIC6400240 × 136 × 20032
TI – 812 WIC8000200 × 200 × 20024
TI – 813 WIC9600240 × 200 × 20032
TI – 814 WIC15000240 × 250 × 25048
TI – 815 WIC18000240 × 250 × 30048

Looking for Exclusive Product ?